Download PDF
Your contact: +49 - 212 8804-0 • info@reo.de

REOMED 600

Isolation transformers

REOMED 600

Unique Selling Point


• certificated according to 3rd. Edition and NRTL
• Optional shield winding between primary and secondary current
• Variants with filter and surge protection
• High efficiency with a very low magnetic field
• Wide range of options
• Small weight
• Short-circuit and overload protection
• Integral start-up current limiter
• Solid aluminum casing
• Equi-potential earthing pin DIN 42801
• Output sockets EN 60320
• EMC tested to EN 60601-1-2

Description


The TÜV-certified REOMED isolating trans-formers are proven and reliable equipment for use with all electrical systems in a medical environment. They reduce the leakage current and thus help to ensure the safety of patients.
REOMED isolating transformers are character-ized by their very low magnetic stray-fields and reliability, whilst also providing high efficiency and easy connectivity.

Application area:

  • Medical systems

Technical Data


  • Rated power : 600 VA
  • Input voltage : 115 - 230 V
Your contact: +49 - 212 8804-0 • info@reo.de

REOMED 600

Isolation transformers

REOMED

600

Nominal input power [VA]

600

600

600

600

600

600

Input voltage [V]

115/230

230

115

115

230

115/230

Output voltage [V]

230

230

230

115

115

115/230

Mains frequency [Hz]

50/60

50/60

50/60

50/60

50/60

50/60

Input power plug

1 piece IE C/EN 60320 (C14) UL 498, CSA C22.2 no. 42

Output power plug

6 piece IEC/EN60320-2-2 (C13) UL 498 CSA C22.2 no. 42

Power switch [0-1]

yes

yes

yes

yes

yes

yes

Input fuse [A]

T6,3/T3,15

T3, 15

T6, 3

T6, 3

T3, 15

T6,3/T3,15

Thermal overload protection [°C]

120

120

120

120

120

120

Max. output current [A]

2,60

2,60

2,60

5,20

5,20

5,20/2,60

Output fuse [A]

-

-

-

-

-

-

Start-up current limiter

yes

yes

yes

yes

yes

yes

Standard

NTC/Timer

NTC/Timer

NTC/Timer

NTC/Timer

electronic

electronic

Earth leakage current 127/254V [µA]         

<300/500

<300/500

<300/500

<300/500

<300/500

<300/500

Insulation / protection resistance [Ω]

>2M/<0,1

>2M/<0,1

>2M/<0,1

>2M/<0,1

>2M/<0,1

>2M/<0,1

Test voltage

primary-secondary.: 4kV AC; primary hosuing / secondary housing: 1,5kV AC

Protection class/med. class/Type

I/I

I/I

I/I

I/I

I/I

I/I

Protection IP

IP20

IP20

IP20

IP20

IP20

IP20

Ambient temperature [°C]

±0 - +40

±0 - +40

±0 - +40

±0 - +40

±0 - +40

±0 - +40

Ambient air humidity [%]

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

 

           
Your contact: +49 - 212 8804-0 • info@reo.de

REOMED 600

Isolation transformers

Type

H (mm)

B (mm)

T (mm)

Weight (kg)

REOMED 600

95 (100*)

170

270

7,5

Your contact: +49 - 212 8804-0 • info@reo.de

REOMED 600

Isolation transformers

REOMED 600

         

Max. altitude [m/feet]

3000/9864

3000/9864

3000/9864

3000/9864

3000/9864

Equi-potential earthing (DIN42801)

POAG - S6/15

Housing, Surface

Powder-coated metal - RAL 7035

Dimensions (BxHxT) [mm/in]

170x95*x270 (*with rubber foot 100)/6.70x3.34*x10.63 (*with rubber foot 3.54)

Weight [kg/lbs]

7,5/16.5

Approval / principles  /
Conformity CE

EN60601-1/:2006-3rd; EN60601-1-2:2007; IEC 60601-1:2005;
IEC 60601-1-2:2007 ;  TÜV-SÜD - NRTL C + US    

Accessories (Options)

Power line

Country-specific

Trigger safety

Mounting kit (1x trigger guard, 4x zinc plated plastic head slotted drews)

Mounting rails

Mounting kit (2x mounting rails, 4x hexagon screw)

Your contact: +49 - 212 8804-0 • info@reo.de

REOMED 600

Isolation transformers

Type

H (mm)

B (mm)

T (mm)

Weight (kg)

REOMED 600

95 (100*)

170

270

7,5