Download PDF
Your contact: +49 - 212 8804-0 • info@reo.de

REOMED II 1580

Isolation transformers

REOMED II 1580

Unique Selling Point


  • Integrated machines
  • Good heat dissipation
  • Compact dimensions
  • Solid aluminium casing
  • Green illuminated mains switch
  • Low total weight
  • Prepared for IEC60601-1:2005 (3rd. Edition)
  • High efficiency with a very low magnetic field
  • Low noise
  • Easy mounting
  • Very low non-load current
  • Low voltage variation
  • Compact design
  • Connection (PTC) for external temperature monitoring

Description


The REOMED II isolation transformer provides reliable leakage current reduction in medical applications. Safe isolation on the input side is effected by using high quality materials and first class production methods. The isolation transformers are designed for low internal losses and so achieve very low no-load losses.
Overload and short-circuit protection on the input and output side is effected by an integral circuit breaker.

Application area:

  • Medical systems

Technical Data


  • Rated power : 1580 VA
  • Input voltage : 230 V
Your contact: +49 - 212 8804-0 • info@reo.de

REOMED II 1580

Isolation transformers

REOMED II

1580

Nominal input power [VA]

1580

Input voltage [V]

230

Output voltage [V]

230

Mains frequency [Hz]

50/60

Input power plug

1 piece IE C/EN 60320 (C20) UL 498, CSA C22.2 no. 42

Output power plug

6 piece IEC/EN60320-2-2 (C13) UL 498 CSA C22.2 no. 42

Power switch [0-1]

yes

Circuit breaker [A]

8

Sec. circuit breaker [A]

7

Thermal overload protection

yes

Temperature sensor

yes, in connection with insulation monitor

Start-up current limiter

NTC (option 10) or electronic (option 50)

Max. output current [A]

6,7

Earth leakage current 254V [µA]         

<500

Insulation / protection resistance [Ω]

>2M/<0,1

Test voltage

primary-secondary.: 4kV AC; primary hosuing / secondary housing: 1,5kV AC

Protection class/med. class/Type

I/I

Protection IP

IP20

Ambient temperature [°C]

±0 - +40

Ambient air humidity [%]

Relative 30 - 75

Max. altitude [m]

2000

Equi-potential earthing (DIN42801)

POAG - S6/15

Housing, Surface

Powder-coated metal - RAL 7035

Dimensions (BxHxT) [mm]

235 x 115* x 325 (with rubber foot 122)

Weight [kg]

17,6

Approval / principles  /
Conformity CE

EN60601-1/:2006-3rd; EN60601-1-2:2007; IEC 60601-1:2005;

IEC 60601-1-2:2007 ;  TÜV-SÜD - 2nd

Accessories (Options)

Power line

Country-specific

Trigger safety

Mounting kit (1x trigger guard, 1x screws kit)

Mounting rails

Mounting kit (2x mounting rails, 4x hexagon screw)

Your contact: +49 - 212 8804-0 • info@reo.de

REOMED II 1580

Isolation transformers


Type

H (mm)

B (mm)

T (mm)

Weight (kg)

REOMED II 1580

115 (122*)

235

325

17,6

*with rubber foot