REOMED 800W

Wide range input for the Medical Technology

REOMED 800W

Unique Selling Point


  • certificated according to 3rd. Edition and NRTL
  • Optional shield winding between primary and secondary current
  • Variants with filter and surge protection
  • High efficiency with a very low magnetic field
  • Wide range of options
  • Small weight
  • Short-circuit and overload protection
  • Integral start-up current limiter
  • Solid aluminum casing
  • Equi-potential earthing pin DIN 42801
  • Output sockets EN 60320
  • EMC tested to EN 60601-1-2

Description


he TÜV-certified REOMED isolating trans-formers are proven and reliable equipment for use with all electrical systems in a medical environment. They reduce the leakage current and thus help to ensure the safety of patients.

REOMED isolating transformers are character-ized by their very low magnetic stray-fields and reliability, whilst also providing high efficiency and easy connectivity.

Application area:

  • Medical systems

Technical Data


  • Rated power : 800 VA
  • Input voltage : 100-130/200-250 V

Type

H (mm)

B (mm)

T (mm)

Weightt (kg)

REOMED 800W

100 (105*)

220

300

12,5

 * with rubber foot

Your contact: +49 - 212 8804-0 • info@reo.de

REOMED 800W

Wide range input for the Medical Technology

REOMED

800W

Nominal input power [VA]

800

Input voltage [V]

100-130/200-250

Output voltage [V]

100-130/200-250

Mains frequency [Hz]

50/60

Input power plug

1 piece IE C/EN 60320 (C14) UL 498, CSA C22.2 no. 42

Output power plug

9 piece IEC/EN60320-2-2 (C13) UL 498 CSA C22.2 no. 42

Power switch [0-1]

yes

Input fuse [A]

T8/T4

Thermal overload protection [°C]

120

Max. output current [A]

8,00-6,16 / 4,00-3,20

Output fuse [A]

-

Start-up current limiter

yes

Standard

electronic

Earth leakage current 127/254V [µA]         

<300/500

Insulation / protection resistance [Ω]

>2M/<0,1

Test voltage

primary-secondary.: 4kV AC; primary hosuing / secondary housing: 1,5kV AC

Protection class/med. class/Type

I/I

Protection IP

IP20

Ambient temperature [°C]

±0 - +40

Ambient air humidity [%]

Relative 30 - 75

Max. altitude [m/feet]

3000/9864

Equi-potential earthing (DIN42801)

POAG - S6/15

Housing, Surface

Powder-coated metal - RAL 7035

Dimensions (BxHxT) [mm/in]

220x100*x300 (*with rubber foot 105)/8.66x3.94*x11.81 (*with rubber foot 4.13)

Weight [kg/lbs]

12,5/27.5

Approval / principles  /
Conformity CE

EN60601-1/:2006-3rd; EN60601-1-2:2007; IEC 60601-1:2005;
IEC 60601-1-2:2007 ;  TÜV-SÜD - NRTL C + US    

Accessories (Options)

Power line

Country-specific

Trigger safety

Mounting kit (1x trigger guard, 4x zinc plated plastic head slotted drews)

Mounting rails